An amperometric biosensor for uric acid determination prepared from uricase immobilized in polypyrrole film.

Article Details

Citation

Cete S, Yasar A, Arslan F

An amperometric biosensor for uric acid determination prepared from uricase immobilized in polypyrrole film.

Artif Cells Blood Substit Immobil Biotechnol. 2006;34(3):367-80.

PubMed ID
16809136 [ View in PubMed
]
Abstract

In order to prepare a biosensor for the determination of uric acid, electropolymerization of pyrrole on Pt surface was carried out with an electrochemical cell containing pyrrole, ferrocene (as a electron mediator) and tetrabutylammonium tetrafluoroborat in acetonitrile by cyclic voltammetry between 0.0 and 1.0 V (vs. Ag/AgCl) at a scan rate of 50 mV/s upon Pt electrode. Uricase was immobilized by a glutaraldehyde/gelatine croslinking procedure on to polypyrrole film after the electropolymerization processes. The response of the biosensor against uric acid was measured after 330 seconds following the application of a constant potential of +0.7 V (vs. Ag/AgCl). The resulting biosensor exhibits excellent electrocatalysis for the uric acid. The amperometric determination is based on the electrochemical detection of H2O2, which is generated in enzymatic reaction of uric acid. The sensor responds to uric acid with a detection limit of 5.0 x 10(-7) M. The sensor remains relatively stable for 5 weeks. Interference effect were investigated on the amperometric response of the biosensor. Determination of uric acid was carried out in the biological fluids by biosensor.