Detection and quantification of affinity ligand leaching and specific antibody fragment concentration within chromatographic fractions using surface plasmon resonance.

Article Details

Citation

Thillaivinayagalingam P, Newcombe AR, O'Donovan K, Francis R, Keshavarz-Moore E

Detection and quantification of affinity ligand leaching and specific antibody fragment concentration within chromatographic fractions using surface plasmon resonance.

Biotechnol Appl Biochem. 2007 Dec;48(Pt 4):179-88.

PubMed ID
17516918 [ View in PubMed
]
Abstract

Rapid analyses of chromatographic steps within a biopharmaceutical manufacturing process are often desirable to evaluate column performance, provide mass balance data and to permit accurate calculations of yields and recoveries. Using SPR (surface plasmon resonance) biosensor (Biacore) technology, we have developed a sandwich immunoassay to quantify polyclonal anti-digoxin Fab fragments used for the production of the FDA (Food and Drug Administration)-approved biotherapeutic DigiFab. The results show that specific Fab may be quantified in all affinity process streams and accurate yield and mass balance data calculated. Control experiments using sheep Fab and Fc indicate that the assay is specific to DigiFab. The quantification of potential leached ligand within chromatographic fractions may also be technically challenging, particularly when low-molecular-mass ligands are covalently coupled with an affinity absorbent. Typical methods to assess ligand leakage such as DDMA (digoxin-dicarboxymethoxylamine; digoxin analogue) often involve the use of labelled ligands and relatively complex and labour-intensive analytical techniques. Using the same analytical methodologies, an assay to detect leached or eluted ligand off the column was developed. The results indicate minimal levels of leached ligand in all chromatographic fractions, with total levels of leached DDMA calculated to be 1.52 microg. This is less than 0.01% of the total amount of DDMA coupled with the laboratory-scale affinity column. The SPR methods described in the present study may be applicable for the rapid in-process analysis of specific polyclonal Fab fragments (within a polyclonal mixture) and to rapidly assess leakage of small molecule ligands covalently attached to chromatographic supports.

DrugBank Data that Cites this Article

Drug Targets
DrugTargetKindOrganismPharmacological ActionActions
Digoxin Immune Fab (Ovine)DigoxinSmall moleculeHumans
Yes
Other/unknown